Security & Threat Intelligence · 06.08.2026, 16:24 UTC
Medixant RadiAnt DICOM
| Schweregrad | info |
|---|---|
| CVE | ↗ |
| Kategorie | Security & Threat Intelligence |
| Quelle | CISA Advisories ↗ |
| Veröffentlicht | 06.08.2026 UTC |
Sicherheitsmeldung mit Schweregrad noch nicht bewertet. Betroffene Kennungen: CVE-2026-17264. Technische Details im Tab „Originaltext“; empfohlene Schritte in der Checkliste.
View CSAF Summary Successful exploitation of this vulnerability could allow an attacker to cause the application to crash if a maliciously crafted DICOM file is opened. The following versions of Medixant RadiAnt DICOM are affected:
RadiAnt DICOM <=2025.2
CVSS Vendor Equipment Vulnerabilities
v3 4.3 Medixant Medixant RadiAnt DICOM Out-of-bounds Write
Background
Critical Infrastructure Sectors: Healthcare and Public Health Countries/Areas Deployed: Worldwide Company Headquarters Location: Poland
Vulnerabilities
Expand All +
CVE-2026-17264
Opening a crafted DICOM file containing malicious JPEG-compressed pixel data triggers an attacker-controlled heap out-of-bounds write, which may allow an attacker to remotely execute arbitrary code. View CVE Details
Affected Products Medixant RadiAnt DICOM
Vendor:Medixant Product Version:Medixant RadiAnt DICOM: <=2025.2 Product Status:known_affected
Remediations MitigationUsers should update to version 2026.1. It is also recommended to open DICOM files only from trusted and reliable sources. Additionally, the application is compiled with exploit mitigation mechanisms enabled, including Control Flow Guard (CFG), Data Execution Prevention (DEP), and Address Space Layout Randomization (ASLR), which significantly reduces the practical exploitability of the issue.https://www.radiantviewer.com/files/RadiAnt-2026.1-Setup.exe
Relevant CWE: CWE-787 Out-of-bounds Write
Metrics
CVSS Version Base Score Base Severity Vector …